Call us at 732 292-1994 or 800 867-6690 | Email us at triadscientific@gmail.com | Visite us on
PLANET BEST
Lab Instrumentation
Lab Equipment
Lab Furniture
Jenway 6270 Fluorometer Filter Fluorometer Jenway 6280 Fluorometer Jenway 6285 Fluorometer Jenway 62 Series Fluorometers Jenway Fluorometer
• High quality optics: pulsed xenon lamp and Total Energy Transfer (TET)
• Intelligent Filter Modules ensure correct filters are fitted for selected method
• Intuitive programming allows fast program creation
• Raw fluorescence or concentration values options
• Standard curve and kinetic functions
• Heated sample holder and sipper pump accessories
• Compatible with DataWay PC software
Three models are available to cover a wide range of applications. The 6280 model is ideal for the most sensitive determinations with emission wavelengths up to 650nm. When higher emission wavelengths are required the 6285 with its red-enhanced detector is the unit of choice. For less sensitive applications and a broader wavelength range up to 1100nm, the low cost 6270 unit is more
suitable.
With press-to-read operation and Total Energy Transfer (TET) technology the output of the high-energy xenon lamp is maximized and its expected life extended so that it should never need replacing within the normal lifetime of the unit. The high quality optics are complemented by the Intelligent Filter Modules (IFM) that are identified by the unit, generating error messages should the wrong filters be fitted or their positions be incorrect.
Each model offers intuitive operation with a user interface based on logical menus that can be navigated from the simple keypad. Up to 20 methods can be created without restriction and saved for future use. They can also be locked against accidental change by password entry. The permanent time and date tag on every stored reading supports Good Laboratory Practice, while calibration reminders and operator identity can also be entered to support conformance and traceability of results.
All models are powered from an external, universal mains adapter suitable for use from 90V to 230V, while the 12V DC input enables use in vehicles or from suitable battery packs.
Features
Model 62 Series Fluorometer
Wavelength selec tion Intelligent Filter Modules (IFM)
Light source Pulsed Xenon lamp with press-to-read operation
Expected life of lamp No replacement required in normal operation
Raw fluorescence readings Yes
Concentration standard curve Blank + up to 6 standards
Kinetics, timed option In both raw fluorescence and concentration modes
Kinetics options 1 reading every 3 to 999 seconds
No. methods stored 20
No. results stored 100 per method
GLP Method lock with password access. Time and date stamped results. Last calibration date and time stored. Calibration interval set. Calibration due reminder. User ID
Outputs RS232
PC software Compatible with DataWay (optional)
Sample handling 10mm square cell holder (standard). Heated sample holder (option). Sipper pump (option)
Power supply 90 to 230V universal power supply (included; 12V DC input to unit)
Size (w x d x h) 365 x 272 x 160mm / 14.5 x 10.8 x 6.4in
Weight 6kg / 13.2lbs
Technical Specification
MODEL 6270
Wavelength range 190 to 1100nm
Sensitivity1 1µg/ml
Dynamic range1 5.2x104
Detector Photodiode
MODEL 6280
Wavelength range 190 to 650nm
Sensitivity1 <1ng/ml
Dynamic range1 5.2x106
Detector PMT2
MODEL 6285
Wavelength range 190 to 850nm
Sensitivity1 <1ng/ml
Dynamic range1 5.2x106
Detector PMT2